Pilbeam's Mechanical Ventilation: Physiological And Clinical Applications, 5th Edition Test Bank
Preview Extract
Chapter 2; How Ventilators Work
Test Bank
MULTIPLE CHOICE
1. The respiratory therapist enters modes and parameters into the ventilator with which of the
following?
a. Control logic
b. Input power
c. User interface
d. Drive mechanism
ANS: C
The user interface or control panel contains certain knobs, dials, or keypads where the
ventilator operator sets or enters certain information to establish how the pressure and
pattern of gas flow is delivered by the machine. Inside the ventilator is the control logic or
control system which interprets the operator settings and produces and regulates the desired
output. The input power is the ventilatorโs power source that provides the energy to enable
the ventilator to perform the work of ventilating the patient. The drive mechanism is a
mechanical device that produces gas flow to the patient.
DIF:
1
REF: pg. 18
2. Which of the following ventilators is pneumatically powered?
a. LTV 1000
b. Bio-Med MVP-10
c. Lifecare PLV-102
d. Intermed Bear 33
ANS: B
The Bio-Med MVP-10 is a fluidic ventilator and uses only gases for its operation. The LTV
1000, Lifecare PLV-102, and Intermed Bear 33 are electrically controlled and powered
ventilators.
DIF:
1
REF: pg. 18
3. A patient being transferred from a hospital to a skilled nursing facility requires mechanical
ventilation with a fractional inspired oxygen (FIO2) of 0.21. The skilled nursing facility has
no piped in gases. Which of the following ventilators will be able to function in the skilled
nursing facility without any extra equipment?
a. Servoi
b. LTV 1000
c. Bird Mark 7
d. Bio-Med MVP-10
ANS: B
The type of ventilator that will be appropriate for this situation is one that is electrically
controlled and powered with a built-in air compressor. The LTV 1000 fits this description.
The Servoi requires both a high-pressure gas source as well as electrical power. The Bird
Mark 7 is a pneumatic ventilator and will not be able to function in this situation. The
Bio-Med MVP-10 is also a pneumatic ventilator that wonโt function in this situation.
DIF:
2
REF: pg. 18
4. The internal circuit of a ventilator allows the gas to go directly from its power source into
the patient. This is known as which of the following?
a. Single-circuit
b. Open loop
c. Closed loop
d. Double-circuit
ANS: A
There are two types of internal circuits, the single- and the double-circuit. The single-circuit
allows the gas to flow from its power input source to the patient. The double-circuit utilizes
a primary power source to generate a gas flow that compresses a mechanism such as a
bellows. The gas within the bellows will then flow to the patient. Open and closed loop refer
to the absence or presence, respectively, of a feedback loop system.
DIF:
1
REF: pg. 21
5. A ventilator for which the primary power source generates a gas flow that compresses
another mechanism and causes the gas from inside the mechanism to be delivered to the
patient is known as which of the following?
a. Single-circuit
b. Double-circuit
c. Closed loop
d. Open loop
ANS: B
In a double circuit ventilator, the primary power source generates a gas flow that
compresses a mechanism such as a bellows or โbag-in-a-chamber.โ The gas in the bellows
or bag then flows to the patient. In a single-circuit ventilator, the primary power source
travels directly to the patient. The closed and open loop refer to whether or not the ventilator
has a feedback loop system.
DIF:
1
REF: pg. 21
6. The function of the exhalation valve is to do which of the following?
a. Adjust the flow going to the patient
b. Close during exhalation to vent patient gas
c. Seal the external circuit during inspiration
d. Determine the volume being delivered
ANS: C
During inspiration, gas fills the balloon and closes a hole in the expiratory valve. Closing of
the hole makes the patient circuit a sealed system. During expiration, the balloon deflates,
the hole opens, and gas from the patient is exhaled into the room through the hole.
DIF:
1
REF: pg. 23
7. In the image, what does โBโ represent?
a.
b.
c.
d.
Expiratory valve line
Exhalation valve
Expiratory line
Main inspiratory line
ANS: C
The external exhalation valve is represented by the letter โCโ in the figure. โBโ is pointing
to the expiratory line. The main inspiratory line is represented by the letter โA.โ The
expiratory valve line is represented by โD.โ
DIF:
1
REF: pg. 19; Figure 2-8A
8. The type of compressors that are used by hospitals to supply wall compressed air has which
of the following?
a. Piston
b. Bellows
c. Rotating blades
d. Moving diaphragm
ANS: A
Hospitals use large, piston-type, water-cooled compressors to supply wall gas outlets.
DIF:
1
REF: pg. 25
9. The power transmission and conversion system of a ventilator is defined as which of the
following?
a. A mechanical device that produces gas flow to the patient
b. An electrical motor that is connected by a special gearing mechanism
c. The system that interprets the settings and produces or regulates the desired output
d. Internal hardware that changes electrical or pneumatic energy into mechanical
energy
ANS: D
The power transmission and conversion system changes the energy from the power source
into mechanical energy. The linear drive piston is a mechanical device that produces gas
flow to the patient. The drive mechanism is an electrical motor that is connected by a special
gearing mechanism. It is the control system that interprets the operator settings and produces
or regulates the desired output.
DIF:
1
REF: pg. 25
10. The volume displacement device that creates a sinusoidal flow waveform is which of the
following?
a. Rotary drive piston
b. Linear drive piston
c. Spring-loaded bellows
d. Proportional solenoid
ANS: A
The rotary drive piston creates a flow pattern that is slow at the beginning of inspiration,
achieves highest speed at mid-inspiration, and tapers off at end-inspiration, creating a
sinusoidal waveform (sine waveform).
DIF:
1
REF: pg. 27
11. Modern intensive care unitsโ (ICU) ventilators regulate gas flow to the patient by using
which of the following?
a. Rotary drive pistons
b. Linear drive pistons
c. Proportional solenoids
d. Spring-loaded bellows
ANS: C
Proportional solenoid valves control flow by opening and closing either completely or in
small increments. These valves, which are driven by various motor-based mechanisms, have
a rapid response time and great flexibility in flow control. The other answers are all volume
displacement devices.
DIF:
1
REF: pg. 25
Document Preview (4 of 257 Pages)
User generated content is uploaded by users for the purposes of learning and should be used following SchloarOn's honor code & terms of service.
You are viewing preview pages of the document. Purchase to get full access instantly.
-37%
Pilbeam's Mechanical Ventilation: Physiological And Clinical Applications, 5th Edition Test Bank
$18.99 $29.99Save:$11.00(37%)
24/7 Live Chat
Instant Download
100% Confidential
Store
Noah Thomas
0 (0 Reviews)
Best Selling
The World Of Customer Service, 3rd Edition Test Bank
$18.99 $29.99Save:$11.00(37%)
Chemistry: Principles And Reactions, 7th Edition Test Bank
$18.99 $29.99Save:$11.00(37%)
Test Bank for Hospitality Facilities Management and Design, 4th Edition
$18.99 $29.99Save:$11.00(37%)
Solution Manual for Designing the User Interface: Strategies for Effective Human-Computer Interaction, 6th Edition
$18.99 $29.99Save:$11.00(37%)
Data Structures and Other Objects Using C++ 4th Edition Solution Manual
$18.99 $29.99Save:$11.00(37%)
2023-2024 ATI Pediatrics Proctored Exam with Answers (139 Solved Questions)
$18.99 $29.99Save:$11.00(37%)